1. Do software programmes which have been developed by outside vendors or academia to support calculations by test facilities using an OECD Test Guideline, and are referenced in OECD Test Guidelines, need to be validated by test facilities?
It is important to note that use by test facilities of the software programmes for these Test Guidelines is voluntary. That is, a test facility is not required to use the programmes to complete a test. However, if a test facility uses such a programme it must have the appropriately scaled validation documentation of risk assessment, user requirement specifications, acceptance testing records, and a listing in the computerised systems inventory.Currently, software programmes have been developed for four OECD Test Guidelines:
While these software programmes were not developed by OECD, they are referenced in the Test Guidelines, and can be downloaded for free on the OECD public website. These software programmes are provided as a courtesy of the developers of the test methods, but have not been validated, reviewed or approved by OECD. The developer, and not the OECD, should be described as the supplier of the software. In addition, the maintenance of the software or of the calculation sheets is not guaranteed by OECD.
If a test facility does use one of these products, and if that facility wishes its test results to be compliant with OECD GLP, it must follow the guidance as described in Advisory Document 17 Application of GLP Principles to Computerised Systems.
These software programmes are considered “Commercial Off-The-Shelf products (COTS)”, as described in OECD Advisory Document 17. Therefore, they require appropriate validation depending on the risk and the complexity of any customisation. If an application (e.g. a spreadsheet) is not complex, it might be sufficient to verify functions against user requirement specifications. In addition, user requirement specifications should be written for any application that is based on a COTS product. Documentation supplied with a COTS product should be verified by test facility management to ensure it is able to fulfill user requirement specifications.